Design and Statistical Validation of Low-Cost Electrogoniometers for Elbow Orthosis Applications
| aut.relation.journal | Biomedical Physics and Engineering Express | |
| dc.contributor.author | Escobedo-Magallán, José F | |
| dc.contributor.author | Téllez-Anguiano, Adriana Del Carmen | |
| dc.contributor.author | Áviles-Jr, Manuel | |
| dc.contributor.author | Garcia, Lorenzo | |
| dc.date.accessioned | 2026-11-08T22:25:55Z | |
| dc.date.issued | 2026-10-30 | |
| dc.description.abstract | This study presents the design, implementation, and statistical validation of four low-cost electromechanical electrogoniometers for elbow joint angle measurement in orthotic applications. The proposed systems are based on distinct sensing principles: an infrared (IR) optical sensor, a calibrated IR optical sensor, a rotary potentiometer, and a slide potentiometer, all integrated into a custom mechanical structure that enables indirect measurement while reducing mechanical stress on sensing components. Experimental validation was performed using a controlled elbow motion simulator and human subjects, with measurements compared against a reference digital goniometer. Performance metrics included mean absolute error (MAE), repeatability, linearity (R²), and statistical agreement. A one-way ANOVA revealed significant differences among sensing modalities (p < 0.05), and Tukey's post hoc analysis confirmed that the IR-based system exhibited significantly higher errors than calibrated IRbased and both potentiometer-based systems. Bland-Altman analysis demonstrated strong agreement for calibrated IR-based and potentiometer-based designs, with minimal bias and narrow limits of agreement. The calibrated IR achieved the highest accuracy (MAE ≈ 2.78°), while the slide potentiometer provided a favorable balance between performance (MAE ≈ 4.72°) and mechanical integration. These results indicate that potentiometer-based electrogoniometers are suitable for wearable rehabilitation systems requiring reliable and realtime joint monitoring. The proposed designs offer cost-effective alternatives for biomechanical instrumentation and support the development of accessible assistive technologies. | |
| dc.identifier.citation | Biomedical Physics and Engineering Express, ISSN: 2057-1976 (Print); 2057-1976 (Online), IOP Publishing. doi: 10.1088/2057-1976/aebbc1 | |
| dc.identifier.doi | 10.1088/2057-1976/aebbc1 | |
| dc.identifier.issn | 2057-1976 | |
| dc.identifier.issn | 2057-1976 | |
| dc.identifier.uri | http://hdl.handle.net/10292/22342 | |
| dc.language | eng | |
| dc.publisher | IOP Publishing | |
| dc.relation.uri | https://iopscience.iop.org/article/10.1088/2057-1976/aebbc1 | |
| dc.rights | © 2026 The Author(s). Published by IOP Publishing Ltd. Open access. | |
| dc.rights.accessrights | OpenAccess | |
| dc.rights.license | Creative Commons Attribution CC BY 4.0 | |
| dc.rights.uri | https://creativecommons.org/licenses/by/4.0/ | |
| dc.subject | Elbow Orthosis | |
| dc.subject | Low-cost electrogoniometers | |
| dc.subject | Medical devices | |
| dc.subject | Elbow Orthosis | |
| dc.subject | Low-cost electrogoniometers | |
| dc.subject | Medical devices | |
| dc.subject | 40 Engineering | |
| dc.subject | 4009 Electronics, Sensors and Digital Hardware | |
| dc.subject | Bioengineering | |
| dc.subject | Physical Rehabilitation | |
| dc.subject | Rehabilitation | |
| dc.subject | 0903 Biomedical Engineering | |
| dc.subject | 1004 Medical Biotechnology | |
| dc.subject | 3206 Medical biotechnology | |
| dc.subject | 4003 Biomedical engineering | |
| dc.title | Design and Statistical Validation of Low-Cost Electrogoniometers for Elbow Orthosis Applications | |
| dc.type | Journal Article | |
| pubs.elements-id | 777132 |